Specification

PARAMETER SPECIFICATION
   Power Supply Single phase AC three-wire system, voltage 180V~250V;

Frequency 50Hz; source resistance ≤ 0.22Ω; Capacity 8kW; max current 37A (instant)

Radiography Power supply voltage: AC 220V ± 22V;
Power frequency: 50Hz ± 0.5Hz;
Power capacity: ≥5kVA;
Maximum allowable internal resistance of power supply: 1Ω
X-ray tube focus Tube voltage: 40-90kV 11 levels, 5kV each; Tube current:

2 levels of 30 and 50mA;

Time: 0.05s-6s in 20 levels.

Mechanical properties When the X-ray generator window is down, the range of focus from the ground is 450mm ~ 1770mm;

The rotation angle of X-ray generator around the horizontal axis is ±180°;

The rotation angle of X-ray generator around the vertical axis is ± 90º;

Overall dimensions 1290(L)×680(W)×1935(H)(mm)

Reviews

There are no reviews yet.

Be the first to review “Mobile X-ray Radiography Equipment AM-NKX-50”

Your email address will not be published. Required fields are marked